Add referance data & x2s5829
This commit is contained in:
136
data/referance/x2s5820/x2s5820.txt
Normal file
136
data/referance/x2s5820/x2s5820.txt
Normal file
@ -0,0 +1,136 @@
|
||||
--------------------------------------------------------------------------------
|
||||
Project.........................................Mars entry non-equilibrium expanding flows
|
||||
Run number................................ x2s5820
|
||||
Date............................................10/10/2024
|
||||
Blame...........................................Mragank, Matt
|
||||
Condition................................... Mars condition, high density, Mars mixture
|
||||
Reservoir..................................... 6.75 MPa
|
||||
Driver............................................83520 Pa 90% He, 92.8 kPa Argon
|
||||
Primary diaphragm.................... 2 mm scored, cold rolled steel 6U374
|
||||
Shock tube................................... 13000 Pa ( Mars mixture)
|
||||
Secondary diaphragm................ 10 micron mylar
|
||||
Acceleration tube....................... 650 Pa lab air (set on new gauge on the bottom of the dump tank)
|
||||
Trig.............................................. Pt5 (changed as wedge is ahead of pitot probes) slot8ai3 on tee piece 8ai4 - cable pt5
|
||||
Lab environment...................... N/A
|
||||
Driver Condition........................x2-lwp-2.0-90-0
|
||||
|
||||
|
||||
|
||||
|
||||
--------------------------------------------------------------------------------
|
||||
EXPERIMENT SETUP
|
||||
|
||||
pitot rake
|
||||
|
||||
|
||||
|
||||
--------------------------------------------------------------------------------
|
||||
EQUIPMENT
|
||||
--------------------------------------------------------------------------------
|
||||
|
||||
1x photodiode (350 - 1100nm)
|
||||
1xShimadzu -- in use
|
||||
1xmini-spec (200-1000nm)
|
||||
Phantom
|
||||
|
||||
|
||||
|
||||
-------------------------- SUSS Spectral Overview ----------------------------
|
||||
--------------------------------------------------------------------------------
|
||||
Dichroic 1 (UV/VIS): DMLP650L
|
||||
Dichroic 2 (VIS/IR): DMLP900L
|
||||
UV FOV: 8mm
|
||||
VIS FOV: 8mm
|
||||
IR FOV: 8mm
|
||||
|
||||
--------------------------------------------------------------------------------
|
||||
---UV---
|
||||
--------------------------------------------------------------------------------
|
||||
Spectrometer: Acton Spectrapro 2300i
|
||||
Centre Wavelength: 430
|
||||
Grating: 150
|
||||
Camera: PIMAX3
|
||||
Gain: 60
|
||||
Delay: 45 us
|
||||
Exposure: 20 us
|
||||
Trigger level: 1 V
|
||||
Slit: 100 um
|
||||
Filter: N/A
|
||||
Trigger: A5 cable from trigger box port 0
|
||||
Monitor: DB1 Cable
|
||||
Phosphor decay delay: default
|
||||
Phsophor decal resolution: default
|
||||
Aperture::12.5 mm
|
||||
|
||||
--------------------------------------------------------------------------------
|
||||
---Visible---
|
||||
--------------------------------------------------------------------------------
|
||||
Spectrometer: Acton Spectrapro 2300i
|
||||
Centre Wavelength: 775
|
||||
Grating: 150
|
||||
Camera: PIMAX4
|
||||
Gain: 50
|
||||
Delay: 45 us
|
||||
Exposure: 20 us
|
||||
Trigger level: 1V
|
||||
Slit: 100 um
|
||||
Filter:
|
||||
Trigger: A3 cable from trigger box port 0
|
||||
Monitor: T1 Cable
|
||||
Aperture: 12.5 mm
|
||||
|
||||
--------------------------------------------------------------------------------
|
||||
---IR---
|
||||
--------------------------------------------------------------------------------
|
||||
Spectrometer: Acton Spectrapro 2500
|
||||
Centre Wavelength: 1300
|
||||
Grating: 75
|
||||
Camera: IRC800
|
||||
Delay: 105us setted in DAQ with 320 counts
|
||||
Exposure: 20 us
|
||||
Slit: 1000 um
|
||||
Filter:
|
||||
Trigger: T2 cable from trigger box port1
|
||||
Monitor: PDTP1G Cable
|
||||
Aperture: 12.5 mm
|
||||
|
||||
--------------------------------------------------------------------------------
|
||||
|
||||
--------------------------------------------------------------------------------
|
||||
Shimadzu
|
||||
--------------------------------------------------------------------------------
|
||||
|
||||
Window : BK7 Diam 209mm top
|
||||
Lens: Nikon AF Micro NIKKOR 200MM with 780-unknown? filter
|
||||
Aperture-Number 16
|
||||
Frame Rates 500.000 kFPS
|
||||
Exposure: 1/2
|
||||
Gain: x2(/x50)
|
||||
Mode: External
|
||||
Trigger @: 0/0 (0-99=100frames)
|
||||
point: 10
|
||||
Delay: 0 microseconds
|
||||
Polarity: Pos
|
||||
Monitor Expose: off
|
||||
Filter:
|
||||
Cables: Trigger: H6 cable from trigger box port 2
|
||||
Monitor: H4 Cable
|
||||
|
||||
--------------------------------------------------------------------------------
|
||||
PHOTODIODE
|
||||
--------------------------------------------------------------------------------
|
||||
N/A
|
||||
|
||||
--------------------------------------------------------------------------------
|
||||
MINI-SPEC (200 - 1000nm)
|
||||
--------------------------------------------------------------------------------
|
||||
N/A
|
||||
|
||||
|
||||
LOG
|
||||
X2s5815
|
||||
Bad shot - very messy - oring ruptured in the secondary may have over pressurised with additional LAB Air.
|
||||
X2s5816
|
||||
Fired at 1.32Pa instead of 13.2pa.....
|
||||
|
||||
|
Reference in New Issue
Block a user